POC / Point-of-care
i15
Edan · China
Specifications
- Model
- i15 Blood Gas and Chemistry Analysis System
- Manufacturer
- Edan Instruments, Inc.
- Measured Parameters
- pH, pCO2, pO2, Na+, K+, Cl-, Ca2+, Glucose, Lactate, Hematocrit (Hct) — up to 10 measured parameters
- pH
- Reportable Range 6.5–8.0, Resolution 0.001, Method: Potentiometric sensor
- pCO2
- Reportable Range 10–150 mmHg, Resolution 0.1, Method: Potentiometric sensor
- pO2
- Reportable Range 10–700 mmHg, Resolution 0.1, Method: Amperometric sensor
- Na+
- Reportable Range 100–180 mmol/L, Resolution 0.1, Method: Potentiometric sensor
- K+
- Reportable Range 2.0–9.0 mmol/L, Resolution 0.01, Method: Potentiometric sensor
- Cl-
- Reportable Range 65–140 mmol/L, Resolution 0.1, Method: Potentiometric sensor
- Ca2+
- Reportable Range 0.25–2.5 mmol/L, Resolution 0.01, Method: Potentiometric sensor
- Glucose
- Reportable Range 20–700 mg/dL / 1.1–38.9 mmol/L, Resolution 1 mg/dL / 0.1 mmol/L, Method: Amperometric sensor
- Lactate
- Reportable Range 2.7–180.2 mg/dL / 0.3–20.0 mmol/L, Resolution 0.1 mg/dL / 0.01 mmol/L, Method: Amperometric sensor
- Hematocrit
- Reportable Range 10–75% / 0.10–0.75 fraction, Resolution 1% / 0.01, Method: Conductance sensor
- Calculated Parameters
- cH+, HCO3-act, HCO3-std, BE(ecf), BE(B), BB(B), ctCO2, sO2(est), Ca++(7.4), AnGap, tHb(est), pO2(A-a), pO2(a/A), RI, pO2/FIO2, and temperature-corrected values: cH+(T), pH(T), pCO2(T), pO2(T), pO2(A-a)(T), pO2(a/A)(T), RI(T), pO2(T)/FIO2, mOsm
- Sample Volume
- 140 µl
- Sample Type
- Whole blood (arterial, capillary, syringe)
- Result Time
- Results within 1 minute of sample aspiration
- Display
- 7-inch colour LCD, 800×480 resolution, touch screen
- Input
- Touchscreen and barcode/QR code scanner
- Battery
- 5000 mAh rechargeable lithium-ion, up to 50 samples per charge (continuous testing)
- Power Supply
- 100–240 V AC, 50/60 Hz
- Dimensions (W×H×D)
- 238×310×153 mm
- Weight
- 3.65 kg
- Operating Temperature
- 10°C to 31°C
- Operating Humidity
- 25% to 80% RH
- Operating Pressure
- 70 to 106.6 kPa
- Patient Record Storage
- up to 10,000 records
- Quality Control
- 3 or 5 level liquid QC; external electronic simulator
- Connectivity
- 4×USB 2.0, RS232, WLAN, LAN
- Built-in Printer
- Yes
- Cartridge configurations
- BG3 (pH, pCO2, pO2), BG4 (pH, pCO2, pO2, Lac), BG8 (pH, pCO2, pO2, Na, K, Cl, Ca, Hct), BG9 (pH, pCO2, pO2, Na, K, Cl, Ca, Glu, Hct), BG10 (pH, pCO2, pO2, Na, K, Cl, Ca, Glu, Lac, Hct), BC4 (Na, K, Cl, Ca, Hct)
- Cartridge storage
- Room temperature with extended shelf life
- Maintenance
- Zero maintenance required (single-use cartridge system)
or email us directly: [email protected]