POC / Point-of-care
e1
Exias Medical · Austria
Specifications
- OPERATIONAL CHARACTERISTICS
- Parameters
- Na+, K+, Cl-, Ca2+, pH, Hct
- Calculated parameters
- nCa2+, tHb(c)
- Min. sample volume
- 20 µL
- Sample type
- whole blood, serum, plasma, urine (undiluted)
- Sample container
- sample tubes, syringes, capillaries (no adapter required)
- Time to result
- 25 s
- Throughput
- TYP 104 samples / hour
- Wet-up time
- 30 min (new cartridge installation)
- ALL-IN-ONE CARTRIDGE
- Number of samples per cartridge
- 150, 300, or 600
- On-board stability
- 28 days / 42 days
- Cartridge storage temperature
- +2 °C to +25 °C
- Cartridge shelf life
- 9 months
- QUALITY CONTROL
- Integrated on-board QC (oQC)
- 3 levels
- oQC measurements / Level
- 30, 50, 60
- Manual QC
- 3 levels
- QC-Trend visualization
- Levey-Jennings-Chart
- Proficiency testing / External Quality Assurance: 3rd party external quality control measurements possible
- CALIBRATION DATA
- Calibration
- Automatic 2-point cal
- Duration
- 30 s
- Interval
- 4, 8, 12 h (configurable)
- INTERFACE
- Touch Display
- 7" TFT
- Resolution
- 800 x 480 px
- Ethernet
- RJ45
- USB flash drive
- 2 USB ports (2.0)
- Barcode reader
- optional
- WiFi dongle
- optional
- STORAGE CAPACITY
- Patient results
- 500,000
- QC & Calibration data
- min. 24 months
- Operators
- 100
- COMMUNICATION
- Data exchange
- LIS
- Protocol
- LIS2-A2 protocol (ASTM)
- Remote access
- Screen sharing
- ADDITIONAL INFORMATION
- Printer
- Built-in 2" thermal printer (easy load)
- Dimensions (W x H x D)
- 20 x 31 x 28 cm
- Instrument weight
- 4.0 kg
- In-use weight
- 4.7 kg
- POWER REQUIREMENTS
- Voltage
- 100 – 240 V
- Power
- mean 15 W; max. 60 W
- Frequency
- 50 / 60 Hz
- ENVIRONMENTAL REQUIREMENTS
- Temperature
- +15 °C to +32 °C
- Humidity
- 15 % to 90 %
- Barometric pressure
- 462 mmHg to 769 mmHg
- MEASURED PARAMETERS (Repeatability within-run precision)
- Na+
- B/S/P 85–200 mmol/l, CV 0.18%; U 5.0–300 mmol/l, CV 0.50%
- K+
- B/S/P 1.0–15.0 mmol/l, CV 0.46%; U 5.0–120 mmol/l, CV 0.73%
- Cl-
- B/S/P 60–150 mmol/l, CV 0.43%; U 5.0–300 mmol/l, CV 0.62%
- iCa2+
- B/S/P 0.1–3.2 mmol/l, CV 0.80%
- pH
- B/S/P 6.5–8.0, CV 0.04%
- Hct
- B 10–75 %, CV 1.64%