Biosystems A15S — IVD analyser distributed by SIRAMED
Occult blood & calprotectin

A15S

Biosystems · Spain

Throughput
150 t/h
Light source
Halogen lamp (6V, 10W)
Lightpath
6 mm
Wavelengths
340 - 405 - 505 - 535 - 560 - 600 - 635 - 670 nm (+1 additional filter can be added by user)
Wavelength accuracy
±2 nm
Spectral range
340 - 900 nm
Photometric range
-0.05 to 3.0 A
Internal resolution
<0.0001 A
Baseline stability
Max. 0.004 A, 30 minutes at 505 nm
Photometric detection system
Silicon photodiode
Maximum capacity of samples
72 (or 48 in A15C)
Barcode reader
External
Size of primary tubes
Ø 13 mm or 15 mm (max. height 100 mm)
Sample well diameter
13 mm
Sample types
Serum, plasma, urine, whole blood, cerebrospinal liquid, semen and biological fluids
Dispensing pump
Ceramic pump of high durability
Dispensing tip
Stainless Steel 110 mm
Level detection
Capacitative
Sample pipetting volume
From 2 μL to 80 μL
Pipetting resolution
0.1 μL
Predilution ratio
From 1:2 to 1:200
Clot detector
No
Tip wash
Inside and outside
Positions for racks (reagents or samples): 4 (reagents or samples) or 2 in A15C
Rack samples capacity
24 sample positions/rack
Number of reagents per reagent rack
10 bottles of 20 or 50 mL
Volume of reagent bottles
20 mL, 50 mL
Cooled reagent
Yes in A15C. 20 reagents max.
Temperature range of refrigerator
10°C below room temperature (at 25°C)
R1 volume
10 μL to 550 μL
R2 volume
10 μL to 200 μL
Dispensing mode (reagent)
Ceramic pump without maintenance
Pipetting resolution (reagent)
1 μL
Tip wash (reagent)
Inside and outside
Reaction volume range
From 180 μL to 800 μL
Number of wells
120
Well material
UV methacrylate
Type of incubation
Dry without maintenance
Temperature
37.0 °C
Temperature accuracy
±0.2 °C
Connectivity to LIS
Yes
IVD - CE
Regulation (EU) 2017/746
Size (w., d., h.)
815 x 665 x 600 mm
Weight
45 kg
Mains voltage
115 to 230 V
Mains frequency
50 or 60 Hz
Electric power
150 VA (200 VA A15C)
Ambient temperature
From 10 to 35°C
Relative humidity
<75%
Altitude
<2500 m
System liquid solution bottle volume
3 L
Washing solution bottle volume
3 L
Waste solution bottle volume
3 L
Operating system
Windows® 10 64 bit (x64)
CPU
Equivalent to Intel Core i3 @3.10 GHz or higher
RAM
512 MB
Hard disk
20 GB or higher
Monitor minimum resolution
800x600
Connector of serial channel
USB
A15S key feature
Cap piercing function to reduce tube manipulation; dedicated software for fecal sample analysis
Fecal calprotectin (FCAL) method
Immunoturbidimetric latex; Analysis mode: Fixed time bireagent; Detection limit: 7.93 µg/g; Linearity limit: 2000 µg/g; Wavelength: 535 nm; On board stability: 30 days; Repeatability: 2.3% at 115 µg/g; Reproducibility: 4.0% at 115 µg/g; Sample type: Feces collected with Stool collection tube
FOB (fecal occult blood) method
Latex enhanced immunoturbidimetric; Measuring range: 20-1000 ng hHb/mL; LOD: 15 ng hHb/mL; LOQ: 20 ng hHb/mL; Wavelength: 505 nm; Linearity: 0-1000 ng hHb/mL; Sensitivity: 96%; Specificity: >99%
Request a quote

or email us directly: [email protected]

  • Laboratory equipment that works.
  • Reagents available.
  • Service that matters.

SIRAMED is an authorised distributor of in vitro diagnostic (IVD) equipment in Romania: haematology, biochemistry, immunology, electrolyte, coagulation, urinalysis, microbiology, HbA1c and point-of-care analysers, together with reagents, consumables and manufacturer-authorised service for medical laboratories, hospitals and clinics. Based in Otopeni, Ilfov — nationwide delivery and technical support since 1991.