Urinalysis
AUTION MAX AX-4060
Arkray · Japan
Specifications
- Measurement object
- Urine
- Reagent
- AUTION Sticks / Uriflet S
- Measurement items
- GLU, PRO, BIL, URO, pH, BLD, KET, NIT, LEU, CRE, ALB, P/C (Protein/Creatinine ratio), A/C (Albumin/Creatinine ratio), S.G. (specific gravity), turbidity, color tone
- Measurement principle
- Test strip: Dual-wavelength reflectance measurement (single wavelength for BLD); S.G.: Reflection refractometry; Color tone: Light-transmission measurement; Turbidity: Light-scattering measurement
- Measurement wavelength
- 5 LED wavelengths (450, 525, 562, 635 and 760 nm)
- Measurement range
- S.G.: 1.000 to 1.050 (Resolution: 0.001); Color tone: color classification; Turbidity: three levels
- Test strip reaction time
- Approximately 60 seconds
- Processing speed
- Normal: Max 225 samples/hour; Small volume: Max 112 samples/hour
- Required sample volume
- Normal/STAT: Minimum 2 mL; Small volume: Minimum 1 mL
- Sample containers
- Sample tubes (length: 95–110 mm, diameter: 14.0–15.8 mm)
- Sample loading capacity
- One-way transportation: Maximum 50 samples; Loop transportation: Maximum 100 samples
- Test strip storage capacity
- Maximum 200 test strips × two storage compartments
- Warm-up time
- Maximum 4 minutes
- Display
- Large color liquid crystal display (800 × 480 dots)
- Memory capacity
- 2500 normal/STAT/small volume tests; 200 control tests; 50 check tests; 100 trouble entries
- Connectivity
- RS-232C compliant; 10BASE-T Ethernet; selectable transmission speeds 300–19200 bps
- Dimensions
- 530(W) × 530(D) × 550(H) mm (including the sampler)
- Weight
- Main body approximately 40 kg; Sampler approximately 4 kg
- Power requirements
- 100–240 V AC, 50/60 Hz, maximum 150 VA
- Operating environment
- Temperature 10–30°C; Humidity 30–60% RH (no condensation)