Immunology
Atellica IM 1600
Siemens Healthineers · Germany
Specifications
- Throughput
- Up to 440 immunoassay tests per hour
- Sample capacity
- Up to 440 samples
- Measurement principle
- Immunoassay analyzer with chemiluminescence technology using advanced acridinium ester (AE) technology
- Reagent compartment
- 42 primary and 35 ancillary refrigerated positions with continuous mixing
- Onboard assay positions
- Up to 42 immunoassay
- Walkaway time
- Up to 5 hours
- Sample volume
- Minimum 10 µL
- Sample types
- Serum, plasma, amniotic fluid, urine, whole blood (assay-dependent)
- Time to first result
- 18 minutes
- Assay menu
- More than 200 specialized assays
- Sample cup compatibility
- 30 different types of sample cups
- Dimensions (H × W × D)
- 150.0 × 145.3 × 116.7 cm (59.1 × 57.2 × 45.9 in.)
- Weight
- 594.7 kg (1308 lb)
- Power consumption
- 2.9 kW maximum
- Power source
- 4.4 kVA (U.S.) / 3.7 kVA (EU), single-phase, 2-pole, 3-wire
- Water consumption
- 6 liters/hour maximum
- Operating temperature
- 18–30°C
- Operating humidity
- 20–80% noncondensing
- Sound pressure level
- 65 dBA average
- Configuration
- Floor model; connects to up to 10 components including Aptio Automation or Atellica Sample Handler
- Quality control
- Onboard automated QC schedule; QC stays onboard and runs automatically
- STAT capability
- STAT samples aspirated within 60 seconds